Tissue array technology for testing interlaboratory and interobserver reproducibility of immunohistochemical estrogen receptor analysis in a large multicenter trial.

von Wasielewski R, Mengel M, Wiese B, Rüdiger T, Müller-Hermelink HK, Kreipe H

Publication: Am J Clin Pathol (2002) • Vol 118 • Page 675-82PMID: 12428786View on PubMed

Purpose

This paper performed a comparative assessment of estrogen receptor (ER) immunohistochemistry among (1) different fixation parameters (fixative and duration, pre-fixation delay), and (2) multiple institutions using tissue microarrays containing negative, low, moderate, and high ER expressing specimens. Confounding sources of variations were also explored.

Conclusion

Optimal ER immunostaining was observed among specimens fixed in neutral buffered formalin for 24 h, with robust reductions observed in specimens subjected to cryosectioning pre-fixation, a pre-fixation delay at 4 degrees C, or fixation in ethanol. Agreement among institutions concerning ER classification varied in relation to the magnitude of ER expression among tumor specimens, with low and intermediate ER expressing tumor specimens generating the most discordance, 56 and 72% agreement, respectively. The authors hypothesize that differences in antigen retrieval protocols may be a confounding variable among institutions.

Studies (2)

Purpose

The purpose of this study was to assess the influence of different fixatives, fixation durations, and a pre-fixation delay on quantifiable ER immunoreactivity in tissue microarrays. Tissue microarray specimens included breast, stomach, lymph nodes, and skin.

Summary of Findings

Optimal ER immunostaining, defined by the percentage of positive nuclei and staining intensity, was observed after immediate fixation in neutral buffered formalin for 24 h, with significant reductions (45-60% that of optimally fixed specimens) observed among specimens post-fixed after cryosectioning, storage at 4 degrees C overnight prior to fixation, or ethanol fixation; conversely, fixation in buffered or unbuffered formalin for as little as 6 h or as long as 72 h resulted in modest reductions in immunostaining (80-100% that of optimally fixed specimens).

  • Biospecimens

    TissueBreast
    TissueStomach
    TissueLymph Node
    TissueSkin
  • Preservation Types

    FormalinEthanol
  • Diagnoses

    NeoplasticCarcinoma
    NeoplasticLymphoma
    NeoplasticSarcoma
    NeoplasticMelanoma
  • Platforms

    ProteinTissue microarray
    ProteinImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionCold ischemia time
    Fixed immediately (unspecified time)
    Overnight at 4 degrees C
    Biospecimen PreservationFixative pH
    5
    7
    Biospecimen PreservationTime in fixative
    168 h
    24 h
    384 h
    6 h
    72 h
    96 h
    Biospecimen PreservationType of fixation/preservation
    Ethanol
    Formalin (buffered)
    Formalin (unbuffered)