Comparison of p53 immunoreactivity in fresh-cut versus stored slides with and without microwave heating.

Shin HJ, Kalapurakal SK, Lee JJ, Ro JY, Hong WK, Lee JS

Publication: Mod Pathol (1997) • Vol 10 • Page 224-30PMID: 9071730View on PubMed

Purpose

The purpose of this paper was to determine if p53 antigenicity is influenced by the duration of room temperature storage of paraffin blocks or slides with affixed formalin-fixed paraffin-embedded sections. The effectiveness of a microwave oven antigen retrieval technique was also investigated.

Conclusion

While p53 status was not influenced by the variables investigated, both the percentage of immuno-positive cells and staining intensity were influenced by the duration of slide storage but not the age of the paraffin block.

Studies (1)

Purpose

The purpose of this study was to determine if p53 antigenicity is influenced by the duration of room temperature storage of paraffin blocks or slides with affixed formalin-fixed paraffin-embedded sections. The effectiveness of a microwave oven antigen retrieval technique was also investigated.

Summary of Findings

p53 status (positive versus negative) was not influenced by the age of the paraffin block used for analysis, the duration of slide storage, or the application of microwave-mediated antigen retrieval. A small albeit significant reduction in the percentage of immuno-positive cells (<6% mean difference) was observed among stored slides in reference to freshly sectioned controls with and without antigen retrieval, when respective controls were compared. A significant reduction in staining intensity was also observed among stored slides compared to freshly sectioned controls, although intensity increased in both stored and fresh slides with antigen retrieval was employed. The time line associated with antigenicity loss was not explored.
  • Biospecimens

    TissueHead and Neck
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    ProteinImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    4-25 y (block storage)
    6-48 months (slide storage)
    Fresh (unspecified duration; slide storage)
    Analyte Extraction and PurificationAntigen retrieval
    Microwave heating in 1% zinc sulfate
    Untreated