Assessment of problems in diagnostic and research immunohistochemistry associated with epitope instability in stored paraffin sections.

van den Broek LJ, van de Vijver MJ

Publication: Appl Immunohistochem Mol Morphol (2000) • Vol 8 • Page 316-21PMID: 11127924View on PubMed

Purpose

This paper investigated the influence of slide adhesives and storage duration (0 d-1 y) and temperature (4-37 degrees C) on the immunohistochemical analysis of formalin-fixed paraffin-embedded (FFPE) tissue sections.

Conclusion

Antigen stability was inversely proportional to the storage duration and temperature of slide-mounted tissue sections, with altered staining observed after as little as 15 d. Both antigen- and staining pattern- specific results were observed. Both heat- and enzyme-mediated antigen retrieval adversely impacted immunostaining. The use of slide adhesives and monoclonal versus polyclonal antibodies did not impact quantifiable immunoreactivity.

Studies (2)

Purpose

The purpose of this study was to determine if the duration and temperature of slide-mounted tissue section storage affects the immunoreactivity of more than 30 antigens. Pituitary glands procured from surgical excision and autopsy were analyzed.

Summary of Findings

Although effects were antigen-specific, the intensity of immunostaining was inversely proportional to storage duration and temperature of slide-mounted tissue sections for several of the antigens evaluated. Specifically, the percentage of stable antigens decreased from 87 to 65 to 50% in sections stored at 4, 21, and 37 degrees C. The maximum storage duration for each antigen was outlined in detail for each storage temperature. Immunostaining intensity was compromised as soon as 1 d after sectioning or remained stable for 360 d. While no differences in immunostaining were observed when monoclonal versus polyclonal antibodies were used with stored versus fresh tissue sections, staining patterns displayed differential results, as membrane staining was more susceptible to prolonged storage durations and elevated temperatures compared to nuclear and cytoplasmic staining. Antigen retrieval (heat- or enzyme-mediated) further impaired immunoreactivity at all storage temperatures.

  • Biospecimens

    TissuePituitary Gland
  • Preservation Types

    Formalin
  • Diagnoses

    Not specified
    Autopsy
  • Platforms

    ProteinImmunohistochemistry
    PeptideImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    0 d
    120 d
    15 d
    180 d
    240 d
    30 d
    300 d
    360 d
    60 d
    90 d
    StorageStorage temperature
    21 degrees C
    37 degrees C
    4 degrees C
    Analyte Extraction and PurificationAntigen retrieval
    0.1% Trysin in CaCl
    0.4% Pepsin in HCl
    Citrate buffer, pH 6 (10 mmol/L)
    None