Discordance between viral loads determined by Roche COBAS AMPLICOR human immunodeficiency virus type 1 monitor (version 1.5) Standard and ultrasensitive assays caused by freezing patient plasma in centrifuged becton-dickinson vacutainer brand plasma preparation tubes.
Salimnia H, Moore EC, Crane LR, Macarthur RD, Fairfax MR
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to assess the magnitude of discordance among plasma HIV-1 viral loads measured by the Bayer branch DNA (bDNA) assay, and Roche COBAS Amplicor Standard and Ultrasensitive assays. Potential sources of variation investigated included viral load range and patient age.
Summary of Findings
Comparisons among Roche COBAS AMPLICOR HIV-1 Standard and Ultrasensitive assays revealed an 18% discordance among the 1,265 case-matched plasma specimens analyzed, with a higher incidence observed among pediatric patients, and specimens with lower Ultrasensitive viral load values (<1000 RNA copy number/ml). bDNA results for HIV-1 viral load displayed a 87% concordance with Ultrasensitive assay results for the 56 case-matched specimens analyzed.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
AIDS/HIV-relatedPlatforms
RNAReal-time qRT-PCRRNABranch DNA assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient age Adult (18-77 y)Pediatric (3 mon-17 y)Real-time qRT-PCR Specific Technology platform Bayer bDNA viral load assayRoche COBAS AMPLICOR HIV-1 Standard AssayRoche COBAS AMPLICOR HIV-1 Ultrasensitive Assay
Purpose
The purpose of this study was to determine the origin of discordance among plasma HIV-1 viral loads measured by the Bayer branch DNA (bDNA)assay, and Roche COBAS Amplicor Standard and Ultrasensitive assays. Potential sources of variation investigated included the type of specimen collection tube, and the type of container used for frozen specimen storage.
Summary of Findings
HIV-1 viral load was concordant for all three technology platforms among plasma specimens collected in EDTA tubes, and those collected in PPT tubes and transferred prior to freezing. However, 62% of plasma specimens collected and frozen in PPT tubes displayed discordance among assays; specifically, elevations in HIV-1 viral load was observed for Standard and Ultrasensitive assay generated values.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
AIDS/HIV-relatedPlatforms
RNAReal-time qRT-PCRRNABranch DNA assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTAPlasma preparation tubes (PPT)Storage Type of storage container New tube (type not specified)Plasma preparation tubes (PPT)Real-time qRT-PCR Specific Technology platform Bayer bDNA viral load assayRoche COBAS AMPLICOR HIV-1 Standard AssayRoche COBAS AMPLICOR HIV-1 Ultrasensitive Assay