The stability of serum hepatitis C viral RNA in various handling and storage conditions.
Fong TL, Charboneau F, Valinluck B, Govindarajan S
Purpose
The purpose of this paper was to determine if serum storage at room temperature or 4 degrees C, or freeze-thaw cycling impacts RT-PCR detection of hepatitis C viral RNA.
Conclusion
Hepatitis C viral RNA was stable in serum specimens stored for 5 days at room temperature or 4 degrees C, as well as those subjected to 5 freeze-thaw cycles.
Studies (1)
Purpose
The purpose of this study was to determine if serum storage at room temperature or 4 degrees C, or freeze-thaw cycling impacts RT-PCR detection of hepatitis C viral RNA. Specimens were centrifuged and separated prior to storage.
Summary of Findings
Detection of hepatitis C viral RNA by nested RT-PCR was not influenced by 5 days of serum storage at room temperature or 4 degrees C or 5 freeze-thaw cycles.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
HepatitisPlatforms
RNART-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Freeze/thaw cycling 1 cycle5 cyclesStorage Storage duration 0 d5 dStorage Storage temperature 4 degrees CRoom temperature