Effect of duration and temperature of storage on serum analyte stability: examination of 14 selected radioimmunoassay procedures.
Kubasik NP, Ricotta M, Hunter T, Sine HE
Purpose
The purpose of this paper was to identify analyte-specific storage temperature and duration thresholds for serum specimens analyzed by radioimmunoassay. Fourteen analytes were assessed.
Conclusion
Insulin was unstable in serum specimens stored at room temperature for 7 days or longer. Gastrin was only stable for 7 days or longer when serum specimens were stored at -70 degrees C. All other analytes failed to differ by more than 10% and were classified as stable at all tested temperatures for up to 14 days.
Studies (1)
Purpose
The purpose of this study was to identify analyte-specific storage temperature and duration thresholds for serum specimens analyzed by radioimmunoassay. Fourteen analytes were assessed in serum separated specimens stored at different temperatures for 1 to 14 days.
Summary of Findings
The following analytes did not vary by more than 10% and as such were stable at all temperatures analyzed for up to 14 days: chlylglycine, cortisol, digoxin, ferritin, follitropin, IgE, lutropin, T4, prolcatin, T3, thyroxin. Insulin and gastrin were unstable at 20 degrees C and -6, 4, 22 degrees C, respectively, after 7 days or longer. Gastrin was stable for up to 14 days when serum specimens were stored at -70 degrees C.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinRadioimmunoassayPeptideRadioimmunoassaySteroidRadioimmunoassayGlycoproteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 1 d14 d7 dStorage Storage temperature -6 degrees C-70 degrees C22 degrees C4 degrees C