Validity for epidemiological studies of long-term cryoconservation of steroid and protein hormones in serum and plasma.
Bolelli G, Muti P, Micheli A, Sciajno R, Franceschetti F, Krogh V, Pisani P, Berrino F
Purpose
The purpose of this paper was to assess the stability of protein (prolactin) and steroid (estradiol, testosterone, progesterone) hormones in serum and plasma after long term frozen storage (1-3 yr) at -80 degrees C.
Conclusion
Serum estradiol and total testosterone proved to be stable for at least 3 years of storage at -80 degrees C, whereas plasma concentrations displayed greater variability with increasing storage durations. Serum free testosterone, however, increased after just 1 year of storage. Serum and plasma prolactin levels were stable for up to 3 years of frozen storage, while serum and plasma progesterone levels fluctuated after 1 or 2 years of storage, respectively.
Studies (2)
Purpose
The purpose of this study was to assess the stability of estradiol, testosterone, progesterone, and prolactin hormones in serum after long term frozen storage.
Summary of Findings
Estradiol, total testosterone, and prolactin did not differ significantly among serum specimens stored for 6 months versus 3 years, whereas free testosterone increased over time by as much as 30%. Serum progesterone concentrations decreased as a function of time by up to 40% after 3 years.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
SteroidRadioimmunoassayProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 1 yr1.5 yr2 yr3 yr6 monthsStorage Storage temperature -80 degrees C
Purpose
The purpose of this study was to assess the stability of estradiol, testosterone, progesterone, and prolactin hormones in plasma after long term frozen storage.
Summary of Findings
Plasma estradiol, total testosterone, and progesterone concentrations varied as a function of frozen storage. Plasma estradiol and progesterone concentrations decreased progressively with storage duration. Variability of plasma estradiol and total testosterone levels were greater in specimens stored for 1 year or more. Conversely, plasma prolactin levels remained stable for up to 3 years.
Biospecimens
FluidPlasmaPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
SteroidRadioimmunoassayProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 1 yr2 yr3 yr6 monthsStorage Storage temperature -80 degrees C