Post-collection, pre-measurement variables affecting VEGF levels in urine biospecimens.
Hayward RM, Kirk MJ, Sproull M, Scott T, Smith S, Cooley-Zgela T, Crouse NS, Citrin DE, Camphausen K
Purpose
The purpose of this paper was to determine the effects of (1) a delay in specimen freezing, (2) freeze-thaw cycling, (3) storage container type, and (4) inclusion of urinary sediment on the quantification of urinary vascular endothelial growth factor (VEGF) levels.
Conclusion
Inclusion of urinary sediment increased VEGF levels compared to supernatant alone; the authors recommend specimen centrifugation prior to analysis. A room temperature delay of up to 4 hours prior to specimen freezing did not affect VEGF levels, nor did the type of polypropylene storage container, or freeze-thaw cycling up to 5 times.
Studies (4)
Purpose
The purpose of this study was to evaluate whether the time between urine biospecimen collection and freezing influenced the measurement of VEGF levels via chemiluminescent immunoassay.
Summary of Findings
No differences were observed among matched specimens stored at room temperature for 0, 1, or 4 hours prior to freezing, although a non-significant increase in VEGF levels was observed after a room temperature delay of 24 hours. Authors recommend freezing urine biospecimens within 4 hours of collection.
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
NeoplasticNormalPlatforms
ProteinImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Time at room temperature 0 h1 h24 h4 h
Purpose
The purpose of this study was to evaluate the extent freeze-thaw cycling influences the measurement of VEGF levels via chemiluminescent immunoassay in urine biospecimens.
Summary of Findings
No significant differences were observed in VEGF levels between specimens subjected to 1, 2, 3, 4, or 5 freeze-thaw cycles. Authors recommend exposing urine biospecimens to no more than five freeze-thaw cycles.
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
NeoplasticPlatforms
ProteinImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Freeze/thaw cycling 1 cycle2 cycles3 cycles4 cycles5 cycles
Purpose
The purpose of this study was to evaluate whether the type of polypropylene storage container influences urinary VEGF levels quantified via chemiluminescent immunoassay.
Summary of Findings
No significant differences in VEGF levels were observed among urine specimens stored in 1.5 ml Eppendorf microcentrifuge tubes, 5 ml Corning cryogenic vials, or 5 ml Greiner Bio-One International cryogenic vials.
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
NeoplasticPlatforms
ProteinImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Type of storage container 1.5 ml Eppendorf microcentrifuge tube5 ml Corning cryogenic vials5 ml Greiner Bio-One International cryogenic vials
Purpose
The purpose of this study was to evaluate whether the presence of urinary sediment influenced the measurement of VEGF levels via chemiluminescent immunoassay in urine biospecimens.
Summary of Findings
VEGF levels were higher in samples containing urinary sediment than in supernatant alone (centrifuged specimens). Although these differences were not statistically significant, the authors recommend centrifugation of urinary samples before measurement of VEGF levels.
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components Contains urinary sedimentNo urinary sediment present