Skip to main content

Venous stasis and routine hematologic testing.

Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC

Publication: Clin Lab Haematol (2006) • Vol 28 • Page 332-7PMID: 16999725View on PubMed

Purpose

The purpose of this paper was to determine if venous stasis, generated by a sphygmometer (0-3 min), influences routine hematological analytes.

Conclusion

After 1 min and 3 min of induced venous stasis, 7 and 10 of the 14 hematological parameters investigated differed significantly from no stasis controls, respectively.

Studies (1)

Purpose

The purpose of this study was to determine if venous stasis, generated by a sphygmometer (0-3 min), influences routine hematological analytes.

Summary of Findings

The following parameters were significantly affected by an induced venous stasis of 1 minute: white blood cell count, red blood cell count, hemoglobin, hematocrit, and lymphocyte and monocyte counts. After 3 minutes of induced venous stasis all investigated parameters differed from paired time zero controls with the following exceptions: main cell volume, main hemoglobin content, mean platelet volume, and basophil count. While red blood cell count, hemoglobin, and hematocrit increased with the duration of venous stasis, total white blood cell counts and subpopulations decreased.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBlood pressure
    Venous stasis (0 min)
    Venous stasis (1 min)
    Venous stasis (3 min)