Venous stasis and routine hematologic testing.
Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC
Purpose
The purpose of this paper was to determine if venous stasis, generated by a sphygmometer (0-3 min), influences routine hematological analytes.
Conclusion
After 1 min and 3 min of induced venous stasis, 7 and 10 of the 14 hematological parameters investigated differed significantly from no stasis controls, respectively.
Studies (1)
Purpose
The purpose of this study was to determine if venous stasis, generated by a sphygmometer (0-3 min), influences routine hematological analytes.
Summary of Findings
The following parameters were significantly affected by an induced venous stasis of 1 minute: white blood cell count, red blood cell count, hemoglobin, hematocrit, and lymphocyte and monocyte counts. After 3 minutes of induced venous stasis all investigated parameters differed from paired time zero controls with the following exceptions: main cell volume, main hemoglobin content, mean platelet volume, and basophil count. While red blood cell count, hemoglobin, and hematocrit increased with the duration of venous stasis, total white blood cell counts and subpopulations decreased.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Cell count/volumeHematology/ auto analyzerProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Blood pressure Venous stasis (0 min)Venous stasis (1 min)Venous stasis (3 min)