Plastic versus glass tubes: effects on analytical performance of selected serum and plasma hormone assays.
Preissner CM, Reilly WM, Cyr RC, O'Kane DJ, Singh RJ, Grebe SK
Purpose
Conclusion
Studies (3)
Purpose
This study comparatively assessed hormone (IGF-1, PTH, hCG, insulin, hCG, cortisol, OHPG) and cancer biomarker (CA-125) levels in paired samples collected and stored in plastic and glass tubes at 4 degrees C for 7 days or frozen immediately.
Summary of Findings
Regardless of storage duration, collection tube dependent differences in cortisol, PHPG, hGH, and beta-hCG levels were not observed. Although modest, statistically significant differences among glass and plastic tubes were observed after 7 days of storage at 4 degrees C for insulin and PTH, and IGF-1 levels among Monoject and Griener plastic tubes, respectively, in comparison to glass tubes. CA-125 levels displayed modest but significantly significant differences among all plastic tubes examined, decreasing as a function of storage time, although in comparison to glass tubes, Vacutainer plastic tubes yielded the lease variable results.
Biospecimens
FluidSerumFluidBloodPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinImmunoassayPeptideImmunoassayGlycoproteinImmunoassaySteroidImmunoassaySteroidLC-MS or LC-MS/MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass tubePlastic Greiner tubePlastic Monoject tubePlastic Vacutainer tubeStorage Storage duration 0 d7 dStorage Storage temperature 4 degrees CImmunoassay Specific Targeted peptide/protein CA-125IGF-1InsulinOHPGPTHbeta-hCGhGH
Purpose
The purpose of this study was to compare the degree of hemolysis among samples collected and stored in plastic and glass tubes at 4 degrees C for 7 days or frozen immediately, as hemolysis may adversely influence insulin quantification.
Summary of Findings
Hemolysis, measured as both hemoglobin concentration and the proportion of samples displaying significant hemolysis, was elevated in glass compared to plastic Vacutainer tubes. The authors report that plastic Monoject and Greiner tubes displayed results similar to Vacutainer plastic tubes.
Biospecimens
FluidSerumFluidBloodPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass tubePlastic Greiner tubePlastic Monoject tubePlastic Vacutainer tubeStorage Storage duration 0 d7 dStorage Storage temperature 4 degrees C
Purpose
This study comparatively assessed ACTH levels in paired samples collected and stored in plastic and glass tubes at 4 degrees C for 0, 4, or 24 hours.
Summary of Findings
ACTH levels did not differ among glass, siliconized glass, or plastic Vacutainer tubes, although levels displayed a modest and nonsignificant downward trend with increasing storage duration at 4 degrees C for all tubes analyzed.
Biospecimens
FluidSerumFluidBloodPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
PeptideImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution EDTA glass tubesEDTA plastic Vacutainer tubesSiliconized EDTA glass tubesStorage Storage duration 0 h24 h4 hStorage Storage temperature 4 degrees C