Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to evaluate different antigen retrieval solutions, and microwave and conventional oven heating regimes for the analysis of FFPE specimens.
Summary of Findings
The authors report that the following conditions resulted in optimal immunostaining while minimizing background staining: microwave heating of sections in a saturated lead thiocyanate solution for two five-minute cycles separated by one minute. Microwave heating was superior to heating in a conventional oven particularly for tissue that had been formalin fixed for 24 hours or longer. Lead thiocyanate proved slightly more efficacious than 1% zinc sulfate and significantly better than distilled water. Enzyme pre-digestion in a trypsin solution was not necessary for optimal antigen retrieval. Immunoreactivity of two representative antibodies (vimentin, pan-cytokeratin) was also enhanced in specimens that were stored in formalin for prolonged periods of time (2 wk-2 yr).
Biospecimens
TissueTonsilTissueLungPreservation Types
FormalinDiagnoses
NeoplasticCarcinomaNot specifiedPlatforms
ProteinImmunohistochemistryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Time in fixative 24 h - 2 yrAnalyte Extraction and Purification Antigen retrieval 1% Zinc sulfateDistilled waterSaturated lead thiocyanateAnalyte Extraction and Purification Protein digestion 0.1% Trypsin, 30 min, 37 degrees CNo trypsin incubationAnalyte Extraction and Purification Temperature of heat-induced retrieval Conventional oven heating, 10 minConventional oven heating, 2 X 5 minConventional oven heating, 5 minMicrowave oven heating, 10 minMicrowave oven heating, 2 X 5 minMicrowave oven heating, 5 min
Purpose
The purpose of this study was to evaluate microwave-mediated antigen retrieval in differentially fixed tissues. Two fixatives (formalin, ethanol) and three fixation temperatures were evaluated at a constant fixation duration of 22 h.
Summary of Findings
The deleterious effects on immunostaining of a formalin sensitive antigen observed at higher fixation temperatures (25 and 37 degrees C) were attenuated by microwave-mediated antigen retrieval with lead thiocyanate. Heat-induced antigen retrieval, while extremely effective for FFPE tissues, did not alter immunostaining of tissues preserved in ethanol.
Biospecimens
TissueSkinPreservation Types
FormalinEthanolDiagnoses
NeoplasticMelanomaNot specifiedPlatforms
ProteinImmunohistochemistryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Temperature of fixation/preservation 25 degrees C37 degrees C4 degrees CBiospecimen Preservation Type of fixation/preservation EthanolFormalin (buffered)