Influences of blood sample processing on low-molecular-weight proteome identified by surface-enhanced laser desorption/ionization mass spectrometry.
Banks R E, Stanley A J, Cairns D A, Berrett J H, Clarke P, Thompson D, Selby P J
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine if the type of anticoagulant or the duration of time before centrifugation (0 -24 h at room temperature) impacts the protein spectral profile of plasma specimens.
Summary of Findings
Compared to time zero controls, plasma samples demonstrated differential peaks in an anticoagulant dependent manner after a 4 hour delay at room temperature before processing; plasma specimens containing EDTA, heparin, and citrate displayed variation in 5-10%, 1-6%, and 0-3% of peaks, respectively, depending upon the chip used for analysis. A delay of 24 hours yielded even more robust changes, with 29% of peaks exhibiting either increased or decreased intensity compared to time zero controls.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinSELDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant CitrateEDTAHeparinNoneBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 0 h1 h2 h24 h30 min4 hSELDI-TOF MS Specific Technology platform CM10 chipH50 chipIMAC30-Cu chip
Purpose
The purpose of this study was to determine if the type of collection tube or the duration of clotting time (0 to 24 h at room temperature) impacts the protein spectral profile of serum specimens.
Summary of Findings
Compared to time zero controls, the spectral profile of serum samples collected in activator tubes showed significant changes in 12-36% of peaks after 4 hours dependent upon the chip used for analysis. However, many of these changes were increases in peak intensity and were present as soon as 30 minutes after venipuncture. Tube-dependent effects were limited to 2 peaks displaying a higher intensity in gel tubes, and 2 peaks displaying a lower intensity compared to activator tubes after 30 minutes at room temperature.
Biospecimens
FluidSerumFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinSELDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Gel serum separatorGreiner vacuette activatorPlain tubeBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 0 min1 h2 h30 min4 h