Skip to main content

Antigen retrieval immunohistochemistry under the influence of pH using monoclonal antibodies.

Shi SR, Imam SA, Young L, Cote RJ, Taylor CR

Publication: J Histochem Cytochem (1995) • Vol 43 • Page 193PMID: 7822775View on PubMed

Purpose

The purpose of this paper was to evaluate the effects of different heat-induced antigen-retrieval buffers under different pH conditions on the immunostaining quality of monoclonal antibodies to cytoplasmic antigens (AE1, HMB45, NSE), nuclear antigens (MIB-1, PCNA, ER), and cell surface antigens (MT1, L26, EMA) in formalin-fixed, paraffin-embedded (FFPE) sections.

Conclusion

The influence of pH on immunostaining quality was antigen-specific, with immunostaining of some antigens (1) unaffected by ph, (2) positively affected at either very low or very high pH, or (3) exhibiting a gradual and continuous increase in immunostaining intensity with pH. While no significant buffer dependent effects were observed at a given pH, Tris-HCl yielded superior results at a high pH. The authors conclude that while heat remains a key factor, as heated distilled water also increased immunostaining, pH should also be optimized for a given antigen.

Studies (1)

Purpose

The purpose of this study was to evaluate the effects of different heat-induced antigen-retrieval buffers under different pH conditions on the immunostaining of monoclonal antibodies to cytoplasmic antigens (AE1, HMB45, NSE), nuclear antigens (MIB-1, PCNA, ER), and cell surface antigens (MT1, L26, EMA) in FFPE sections. An unspecified number of tonsil, pancreas, lymph nodes with metastatic squamous cell carcinoma, melanoma, and breast carcinoma were fixed in 10% formalin for 24 h and paraffin embedded. Five µm sections were dried for an hour at 60˚C, dewaxed, rehydrated and endogenous peroxidase was blocked using unspecified methods. Slides were placed in jars containing the 7 different antigen retrieval solutions tested prepared at pH 1-10 depending on solution and microwaved for 5 +5 min at 2450 mHz and 600 W. When required more antigen retrieval solution was added between microwave steps. After microwaving slides remained in antigen retrieval solution for 15 min before washing and immunostaining with the different antibodies. Immunostaining of each slide was scored at extremely strong, strong, moderate, weak or negative by three blinded authors independently..

Summary of Findings

The authors report that pH had a greater influence on immunostaining than buffer solution; no significant differences were seen in maximal staining achieved with each different buffer. The influence of pH fell into three categories: antibodies with no significant variation correlated to pH (including L26, PCNA, AE1, EMA, NSE); antibodies that stained well in the very low and high pH ranges, but stained poorly under moderately acidic conditions (pH 3-6; including MIB-1, ER)); and antibodies for which staining improved as pH increased (MT1, HMB45). They concluded that, although heating is the most important factor in antigen retrieval, pH of the retrieval solution may also play a role and should be optimized for the antibody of interest. The damage to or separation of the tissue that is occasionally noted when heating a strongly alkaline solution was circumvented with Tris-HCl and sodium acetate buffers at a pH of 9, which adequately preserved morphology.

  • Biospecimens

    TissueLymph Node
    TissueTonsil
    TissuePancreas
    TissueSkin
    TissueBreast
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticCarcinoma
    NeoplasticMelanoma
    Not specified
  • Platforms

    ProteinImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentspH
    2.0 - 9.4
    Analyte Extraction and PurificationAntigen retrieval
    Acetate
    Barbitol sodium/sodium acetate
    Citrate
    Dimethyl glutaric acid/NaOH
    Distilled water
    Phosphate
    Sodium phosphate/citric acid
    Tris-HCl