Preanalytical and analytical variation of surface-enhanced laser desorption-ionization time-of-flight mass spectrometry of human serum.
Albrethsen J, Bogebo R, Olsen J, Raskov H, Gemmeltoft S
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine if the protein profiles of serum and plasma are influenced by the following preanalytical factors: patient age, gender, fasting, diurnal rhythm, clotting time, storage temperature.
Summary of Findings
As determined by principal components analysis, clot time at room temperature affected peak intensity, with robust differences observed between 30 and 60 minutes, and small differences from 60 to 120 minutes; while no changes were observed in EDTA-plasma samples stored at room temperature for 30 to 120 minutes. While storage of serum samples at room temperature yielded significant differences in peak intensities after 60 minutes, no effects were observed in samples stored on wet ice for equivalent durations. The authors note no correlation between protein peak variation and the following variables: age, gender, fasting, diurnal rhythm, and serum freeze-thaw cycling.
Biospecimens
FluidSerumFluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinSELDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition FastingNon-fastingPreaquisition Patient age 50 - 80 yearsPreaquisition Patient gender FemaleMaleBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Freeze/thaw cycling 1 cycle2 cycles3 cyclesStorage Storage duration 120 min240 min30 min60 minStorage Storage temperature 4 degrees CRoom temperature
Purpose
Summary of Findings
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinSELDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method Ethanol precipitationImmunoaffinity chromatographySELDI-TOF MS Specific Technology platform CM10H50IMAC30NP20Q10SAX2WCX2